Multiparametric Patient Monitor iM8 (ECG, NIBP, PR, TEMP, RESP, SpO2)
Description
The iM8 multiparameter monitor is designed to monitor the following parameters: ECG, respiration (RESP), SpO2, NIBP, and temperature (TEMP). Thanks to its optimized performance and accurate measurements, the iM8 meets basic monitoring needs. It has been developed with Wi-Fi connectivity so that the healthcare professional can check the patient's health status conveniently from anywhere. Additionally, when connected to EDAN's MFM-CMS software, it is also possible to check the patient's status from any PC or tablet via the web viewer.
Features:
- 12.1" high-resolution color TFT LCD display.
- Pacemaker detection.
- Protection against interference from electrosurgery and defibrillation.
- Alarm tone (pulse tone modulation).
- Unique iSEAP algorithm optimized especially for patients with arrhythmias.
- iMAT SpO2 algorithm with excellent motion resistance and low perfusion resistance performance.
- NIBP with SP10 clinical validation.
- Built-in Li-ion rechargeable battery.
- Nurse call function.
- Suitable for adult, pediatric, and neonatal patients.
Optionally and at an additional cost, the following functionalities can be included: 2-IBP, Respironics CO2, EDAN G2 CO2, thermal printer, WiFi-WLAN kit, and VGA output for second screen or TV.
Technical specifications:
• ECG+PR (HR):
- Lead type: selectable 5/3 leads.
- Gain selection: Auto, x0.125, x0.25, x0.5, x1, x2, x4.
- Sweep speed: 6.25, 12.5, 25, 50mm/s.
- ECG HR (HR) range:
• Adult: 15-300 bpm.
• Pediatric/Neonate: 15-350 bpm.
- Resolution: 1 bpm.
- Accuracy: ±1 bpm or ±1% (whichever is greater).
- Filters:
• Diagnostic mode: 0.05-150 Hz.
• Monitoring mode: 0.5-40 Hz.
• Surgical mode: 1-20 Hz.
- ST segment detection:
• Measurement range: -2.0 mV-2.0 mV.
• Alarm range: -2.0 mV-2.0 mV.
• RESP:
- Method: transthoracic impedance.
- Operating mode: auto/manual.
- Respiratory rate range:
• Adult: 0-120 rpm.
• Neonate/Pediatric: 0-150 rpm.
- Resolution: 1 rpm.
- Apnea alarm: 10, 15, 20, 25, 30, 35, 40 s.
- Bandwidth: 0.2-2.5 Hz (-3dB).
- Sweep speed: 6.25, 12.5, 25, 50 mm/s.
• SpO2:
- Measurement and alarm range: 0-100% (EDAN SpO2).
- Resolution: 1 bpm.
- Accuracy:
• ±2% (70-100%, Adult/Pediatric).
• ±3% (70-100%, Neonate).
- HR measurement and alarm range: 25-300 bpm.
- Update frequency: 1s.
• NIBP (NIBP):
- Method: automatic oscillometric.
- Operating modes: manual/automatic/continuous.
- Auto measurement at time interval: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 240, 480 min.
- Measurement units: mmHg/kPa.
- Measurement types: systolic, diastolic, mean.
- Pressure range:
• Adults:
- Systolic: 40-270 mmHg.
- Diastolic: 10-215 mmHg.
- Mean: 20-235 mmHg.
• Pediatrics:
- Systolic: 40-200 mmHg.
- Diastolic: 10-150 mmHg.
- Mean: 20-165 mmHg.
• Neonates:
- Systolic: 40-135 mmHg.
- Diastolic: 10-100 mmHg.
- Mean: 20-110 mmHg.
- Resolution: 1 mmHg.
- Accuracy:
• Maximum mean error: ±5 mmHg.
• Maximum standard deviation: 8 mmHg.
- HR measurement range from NIBP: 40-240 bpm.
- Resolution: 1 bpm.
- Accuracy: ±3 bpm or 3.5% (whichever is greater).
- SP10 validation:2002.
• TEMP:
- Measurement and alarm range: 0-50°C (32-122°F).
- Resolution: 0.1°C.
- Accuracy: ±0.1°C (without probe).
- Channel: dual channel; provides T1, T2 and ∆T.
• Display:
- Color TFT LCD: 12.1”.
- Resolution:
• 800x600 (iM8/iM8A).
• 800x480 (iM8B).
- Parameters plotted: up to 8.
- Displayed curves: up to 11.
- Sweep speed: 6.25, 12.5, 25, 50 mm/s.
• Power supply:
- AC power: 100-240V AC, 50/60 Hz.
- Battery: rechargeable Li-ion 14.8V.
- Battery life:
• 1.5 hours (2.1 Ah).
• 3 hours (4.2Ah).
• Dimensions:
- Length: 32 cm.
- Depth: 15 cm.
- Height: 26.5 cm.
- Weight: approximately 5 kg.