Reference. 400133776

BM5 Multiparameter Monitor (ECG, NIBP, PR, TEMP, RESP, SpO2)

BM5 multiparameter monitor (ECG, NIBP, PR, TEMP, RESP, SpO2)
Personalized attention
Personal advice
BM5 Multiparameter Monitor (ECG, NIBP, PR, TEMP, RESP, SpO2)
BM5 Multiparameter Monitor (ECG, NIBP, PR, TEMP, RESP, SpO2)
3.100,00€ (3.751,00€ VAT INCLUDED)
Buy
Secure payment
Approximate delivery time of 10 to 20 days

Description

BM5 multiparameter monitor.

Features:
- High-performance ECG analysis. ST level. PVC housing. 13 types of arrhythmia analysis. 7 channels.
- 10.4" TFT LCD color touchscreen, with 6 waveform displays.
- Individual sweep speed adjustment for each parameter.
- Screen freeze enabled/disabled.
- OxyCRG, drug calculation.
- External HDMI output for large screens.
- Direct DC input for vehicles/ambulance.
- Real-time bed-to-bed connectivity, up to 8 monitors without server or central station.
- SpO2 technology with very low perfusion detection.
- Graphical and tabular trend printouts.
- Storage of 168 full hours of trend data.
- Long-life lithium-ion battery.
- Direct HL7 EMR with editable menu.
- USB memory storage.
- Enhanced and expandable multiparameters:
• Main/secondary EtCO2 (optional).
• High-performance Suntech NIBP (BM5ICU).
High-performance ECG detection and analysis:
• Improved input impedance for dry skin or electrode patch.
• ST level, PVC count, real-time display.
• 13 types of arrhythmia analysis.
• Pacemaker detection function.

- Connection to the central monitoring system via LAN or W-LAN (optional).
- Selectable software language (GB, DE, FR, IT, PT, PL, RO, HU, ES, CZ, BG, RU, TR, CN).

Technical Specifications: 
GENERAL 
• Dimensions / Weight: 238x163xh 250 mm/4 kg.
• Display of up to 6 waveforms (ECGx2, SpO2, RR or EtCO2 IBPx2).
• Indicators: 
- Categorized alarms (3 priority levels).
- Visual alarm on handle
- Heart rate 
- SpO2 pulse tone
- Battery status, 
- External power LED 
• Thermal printer: 25, 50 mm/sec (paper width 58 mm).
• Battery: Rechargeable Li-ion (3 hours of uninterrupted operation).
• Interfaces: 
- Digital LAN output for data transfer.
- Connection to the nurse call system.
- Output for defibrillator connection.
• DC input connector: 18 V DC, max 2.8 A. 
• Operating voltage: 100-240 V, 50-60 Hz.

SpO2
• Saturation range: 0% to 100% 
• Pulse rate range: 30 to 254 bpm.
• SpO2 accuracy: ± 2 digits (70% to 100%).
• Pulse rate accuracy: ± 2 bpm.

ECG+PR (RC)
• Cable type: 3 leads, 5- leads
• Lead selection:
- 3 leads: I, II, III.
ECG waveforms:
- 3 leads 1 channel.
- 5 leads: 1/2/7 channels.
Heart rate range:
- Adult: 30-300 bpm.
- Neonate / pediatric: 30-350 bpm.
• Heart rate accuracy: ± 1 bpm or ± 1%, whichever is greater.
• Sweep speed: 6.25, 12.5, 25, 50 mm/sec.
• Filter: 
- Diagnostic mode: 0.05 to 150 Hz.
- Monitoring mode 0.5 to 40 Hz.
- Moderate: 0.5 to 25 Hz.
- Maximum: 5 to 25 Hz.
• ST segment detection range: - 2.0 to 2.0 mV.
• Arrhythmia analysis: Asystole, Vtach, Vfib, Bigeminy, Accvent Couplet, Irregular, Pause, PVC, Ront, Trigeminy, Vbrady, Shortrun.
• Pacemaker detection mode: waveform display indicator (user selectable).
• Protection: against electro-surgical interference and defibrillation.

NIBP (PANI)
• Technique: Oscillometric with linear deflation.
• Measurement modes: Manual-Automatic-Continuous.
• Measurement range:
- Adult pressure: 20 to 260 mmHg.
- Pediatric pressure: 20 to 160 mmHg.
- Neonatal pressure: 20 to 130 mmHg.
- Pulse rate: 30-220 bpm.
• Pressure range: 0 to 300 mmHg ± 3 mmHg.
•  Safety pressure:
Adult: 300 mmHg.
Neonate: 150 mmHg.
• Accuracy: meets the accuracy requirements of ANSI/AAMI ISO 81060-2-2013.

RESP
• Range: 5 to 120 breaths/min ±1 b/min.
• Apnea alarm: yes.

TEMP 
• Range: 0° to 50° C (32°F to 122°F).
• Accuracy: ± 0.1°C (25 to 50°C), ± 0.2°C (0 to 24°C).
• Compatible with: YSI series 400 temperature probes.

IBP (optional) 
• Channels: 2.
• Measurement range: -50 to 300 mmHg.
• Measurement accuracy:
- ± 1 mmHg (<100 mmHg).
- ± 1% of the reading (> 100 mmHg).
• Pulse frequency count range: 0 to 300 beats/min.
• Zero balance:
- Range ± 200 mmHg.
- Accuracy ± 1 mmHg.
- Drift ± 1 mmHg over 24 hours.
• Transducer sensitivity: 5 µV/V mmHg.

EtCO2 (optional) 
• Measurement mode: Sidestream or Mainstream.
• Measurement range: 0 ~ 150 mmHg, 0 to 19% .
• Breathing range: 2 ~ 150 bpm.
• Accuracy: 
- ± 2 mmHg, 0 ~ 40 mmHg.
- ± 5%, 41 ~ 70 mmHg.
- ± 8%, 71 ~ 100 mmHg.
- ± 10%, 101 ~ 150 mmHg.
• Breathing accuracy: ± 1 breath per minute.

Alternative products