Reference. 40432020-5370
12.1" Proview 12 Multiparametric Touch Monitor (EtCO2, ECG, NIBP, IBP, TEMP, RESP, SpO2, CO)
12.1" Proview 12 Multiparametric Touch Monitor (EtCO2, ECG, NIBP, IBP, TEMP, RESP, SpO2, CO)
Description
Proview 12 multiparameter monitor with 12.1" color touch screen designed for accurate measurement of ECG, SpO2, NIBP, respiration, temperature, cardiac output (C.O) and CO2 levels (EtCO2) in both adults and children and neonates. It incorporates a state-of-the-art printer that allows results to be printed in a few seconds, thus speeding up the work pace. It also has a USB connection and a LAN output to transfer data to other electronic devices for further analysis.
Characteristics:
- 27 arrhythmia analysis modes.
- Early Warning Score (EMS): assessment of the urgency of medical action.
- Glasgow Coma Scale (GCS).
- 800x600 pixel resolution screen.
- Monitoring of the respiratory interval in neonates.
- Designed with a space to store accessories at the back.
- Internal rechargeable lithium-ion battery with up to 4 hours of operation with full charge.
- Battery status indicator.
- 7 waveforms for monitoring ECG, SpO2, respiration, EtCO2 and IBP.
- Alarms categorized into 3 priority levels.
- Heartbeat tracking with tone.
- External power LED.
- Available internal languages: English, Chinese, French, German, Italian, Polish, Spanish, Portuguese, Russian, Czech, Turkish, Danish, Dutch, Finnish, Hungarian, Norwegian, Swedish, Romanian, Syrian and Greek.
- 2-year warranty.
Certificates: in accordance with the requirements for electromedical equipment established in UNE-EN 60601-1 and 60601-2 standards.
Technical Specifications:
- Dimensions: 175x320x262 mm
- Weight: 4 kg
- Parameters:
ECG:
- Leads: 3/5/6/12
- Heart rate range:
- Adult: 10 to 300 bpm
- Pediatrics: 10 to 350 bpm
- Heart rate accuracy: ± 1 bpm
- ST segment:
- Detection range: - 2.0mV + 2.0mV
- Accuracy: -0.8mV + 0.8mV
- Resolution: 0.01 mV
- Bandwidth: 0.5 Hz ~ 40 Hz
- Gain: 6.25, 12.5, 25, 50 mm/sec.
- Input impedance: >5MΩ
- CMRR: >100dB
- Signal input range: 10.0mV ~ +10.0mV
- Arrhythmia analysis: VTAC / VFIB / ASYSTOLIC etc.
- Lead-OFF detection: screen indicator
SpO2:
- Measurement range: 0 % - 100 %
- Pulse range: 25 - 300bpm
- SpO2 accuracy
- 70% - 100%: <3%
- 0% - 69% unspecified
- Pulse accuracy: ± 3 bpm
- Perfusion index range: 0.05-20.0%
- Perfusion index accuracy: ± 0.1% or ± 10% of the indicated value, whichever is greater
NIBP (PANI):
- Method: oscillometric
- Operating mode: manual, automatic and STAT
- Automatic measurement interval:
- 1, 2, 2.5, 3, 5, 10, 15, 20, 30 min
- 1, 1.5, 2, 4, 8 hours
- STAT: 5 min
- Systolic range:
- Adult: 30 - 270 mmHg
- Pediatric: 30 - 235 mmHg
- Neonate: 30 - 135 mmHg
- Diastolic range:
- Adult: 10 - 220mmHg
- Pediatric: 10 - 220 mmHg
- Neonate: 10 - 110 mmHg
- Average:
- Adult: 20 - 235 mmHg
- Pediatric: 20 - 235 mmHg
- Neonate: 20 - 125 mmHg
- Pressure accuracy: ± 3 mmHg
- Adjustable cuff inflation pressure.
Respiration:
- Range: 0 - 150 rpm
- Accuracy: ± 2 rpm or ± 2%, whichever is greater
- Resolution: 1 rpm
Power supply:
- Input: 100-240V (50/60 Hz)
- Input power: 100VA
Data memory:
- Alarms: 3,000 groups
- Trend: 180 hours
- NIBP: 2,400 groups
- Holographic curve: 72 hours
- 12.1" color touch screen (800x600 pixels)
- Printer:
- Speed: 12.5, 25, 50 mm/sec.
- Paper width: 50 mm
Temperature:
- Parameter: T1, T2, TD.
- Probe: YSI400 series probe (2252Ω at 25 °C)
- Range: 0.0 °C to 50.0°C (32 °F to 122 °F)
- Accuracy: ± 0.1°C or ± 1°F (probe exclusive)
- Resolution: ± 0.1°C or ± 1°F
IBP (PAI):
- Transducer sensitivity: 5uV/V/ mmHg, ± 2%
- Transducer impedance: 300 Ω to 3000 Ω
- Range: 50 mmHg to 360 mmHg
- Accuracy: ± 2 mmHg or ± 2% of the reading, whichever is greater (excluding the transducer)
- Resolution: 1 mmHg
- Unit: mmHg, kPa, cmH2O
CO:
- Range:
- C.O.: 0.1 L/min to 20 L/min
- TB: 23.00 °C to 43 °C
- TI: -0.1 °C to 27 °C
- Resolution:
- C.O.: 0.1 L/min
- TB: 0.01 °C
- TI: 0.1 °C
- Accuracy:
- C.O.: ± 5% or ± 0.1 L/min, whichever is greater
- TB: ± 0.1 °C
- TI: ± 0.1 °C
EtCO2:
- MicroFlow CO2:
- Range: 0% to 25% (0 mmHg to 190 mmHg)
- Accuracy: ± 0.43% (+8% of reading)
- Resolution: 0.1% or 1 mmHg
- Unit: %, mmHg, kPA.
- Preheating time: <10 s (concentration report and maximum accuracy range)
- Rise time: <3 s (including delay time and rise time)
- Sample flow rate: 50 ± 10 mL/min
- AwRR range: 0 rpm to 150 rpm
- AwRR accuracy: ± 1 rpm
- Conventional CO2:
- Range: 0% to 25% (0 mmHg to 190 mmHg)
- Accuracy: ± 0.43% (+8% of reading)
- Resolution: 0.1% or 1 mmHg
- Unit: %, mmHg, kPa
- Preheating time: <10 s
- Rise time: <90 ms
- AwRR range: 0 rpm to 150 rpm
- AwRR accuracy: ± 1 rpm
Includes:
- Thermal printer.
- Adult size blood pressure measurement cuff.
- 5-lead patient ECG cable.
- Single-use electrodes.
- 3m NIBP cable.
- Adult SpO2 sensor.
- 2m SpO2 extension cable.
- Battery: 2.5 Ah, 4 hours.