Reference. 1208TOO10
Adult defibrillation electrodes for Progetti and General Electric equipment. 1 pair
Non-sterile disposable electrodes designed for Progetti and General Electric defibrillation equipment. Suitable for monitoring with electrocardiogram and synchronized cardioversion.
Description
Non-sterile disposable adult electrodes for defibrillation, electrocardiogram monitoring, and synchronized cardioversion. Made with biocompatible and highly resistant materials, they are capable of withstanding up to 15 consecutive discharges of 360 joules. They can be used with Progetti-brand defibrillation devices, such as the Rescue Life or Rescue Sam models, and the General Electric brand, such as the Responder AED.
General features
- Single-use non-sterile electrodes
- Indicated for defibrillation, ECG monitoring and synchronized cardioversion
- Highly resistant, capable of withstanding up to 15 discharges of 360 J sequentially
- Free of phthalates, natural latex and compounds of biological origin
- Non-cytotoxic, non-irritating and non-sensitizing
Technical specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the date of manufacture
- Compatibility: Progetti (Rescue Life and Rescue Sam) and General Electric (Responder AED) equipment
Materials
- Support: 1 millimeter thick white polyethylene flexible foam and waterproof cell with biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic base hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic PVC
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate / thermoplastic elastomer
Electrical properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS Directive
- Directive 93/42/EEC, class I medical product
General features
- Single-use non-sterile electrodes
- Indicated for defibrillation, ECG monitoring and synchronized cardioversion
- Highly resistant, capable of withstanding up to 15 discharges of 360 J sequentially
- Free of phthalates, natural latex and compounds of biological origin
- Non-cytotoxic, non-irritating and non-sensitizing
Technical specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the date of manufacture
- Compatibility: Progetti (Rescue Life and Rescue Sam) and General Electric (Responder AED) equipment
Materials
- Support: 1 millimeter thick white polyethylene flexible foam and waterproof cell with biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic base hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic PVC
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate / thermoplastic elastomer
Electrical properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS Directive
- Directive 93/42/EEC, class I medical product