Reference. 1208TOO16
Adult defibrillation electrodes for HP/Philips and Agilent Philips equipment. 1 pair
Non-sterile disposable electrodes intended for ECG monitoring, synchronized cardioversion, and defibrillation. Compatible with HP/Philips and Agilent Philips devices.
Description
These disposable electrodes allow defibrillation processes, synchronized cardioversion, and electrocardiogram monitoring in adult patients. Made with toxin-free and allergen-free materials, they are biocompatible, guarantee safe use in all types of patients, and can withstand up to 15 sequential discharges of 360 joules each. To use them, a Heartstart XL defibrillation unit from HP/Philips or a Heartstart FR2+ from Agilent Philips is required.
General features
- Disposable electrodes designed for defibrillation, synchronized cardioversion and ECG monitoring in adults
- Highly resistant, they can withstand up to 25 discharges of 360 J sequentially
- Made with biocompatible materials that do not irritate, sensitize, or cause cytotoxicity
- Free of phthalates, natural latex, and biological materials
Technical specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the manufacturing date
- Compatibility: HP/Philips (Heartstart XL) and Agilent Philips (Heartstart FR2+) equipment
Materials
- Support: 1 millimeter thick white polyethylene foam with waterproof cell and biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic base hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic PVC
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate/thermoplastic elastomer
Electrical properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS regulation
- Directive 93/42/EEC, class I medical product
General features
- Disposable electrodes designed for defibrillation, synchronized cardioversion and ECG monitoring in adults
- Highly resistant, they can withstand up to 25 discharges of 360 J sequentially
- Made with biocompatible materials that do not irritate, sensitize, or cause cytotoxicity
- Free of phthalates, natural latex, and biological materials
Technical specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the manufacturing date
- Compatibility: HP/Philips (Heartstart XL) and Agilent Philips (Heartstart FR2+) equipment
Materials
- Support: 1 millimeter thick white polyethylene foam with waterproof cell and biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic base hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic PVC
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate/thermoplastic elastomer
Electrical properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS regulation
- Directive 93/42/EEC, class I medical product