Reference. 1208TOO07
Adult defibrillation electrodes for Agilent Philips equipment. 1 pair
Disposable electrodes compatible with Agilent Philips equipment. Specially designed for defibrillation, synchronized cardioversion, and ECG monitoring.
Description
Non-sterile, single-use electrodes indicated for synchronized cardioversion, defibrillation, and electrocardiogram monitoring in adult patients. Made from biocompatible and non-toxic materials and designed to withstand up to 15 consecutive discharges of 360 joules. Compatible with Agilent Philips Heartstart XL and Heartstart FR2+ defibrillators.
General features
- Non-sterile disposable electrodes designed for defibrillation, synchronized cardioversion, and ECG monitoring
- Highly resistant, capable of withstanding up to 15 consecutive discharges of 360 J
- Free from natural latex, phthalates, and biologically sourced compounds
- Non-cytotoxic, non-sensitizing, and non-irritating
Technical specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the date of manufacture
- Compatibility: Agilent Philips equipment (Heartstart XL and Heartstart FR2+)
Materials
- Support: 1 millimeter thick white polyethylene foam and waterproof cell with biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic-based hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic PVC
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate/thermoplastic elastomer
Electrical properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS regulations
- Directive 93/42/EEC, Class I medical product
General features
- Non-sterile disposable electrodes designed for defibrillation, synchronized cardioversion, and ECG monitoring
- Highly resistant, capable of withstanding up to 15 consecutive discharges of 360 J
- Free from natural latex, phthalates, and biologically sourced compounds
- Non-cytotoxic, non-sensitizing, and non-irritating
Technical specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the date of manufacture
- Compatibility: Agilent Philips equipment (Heartstart XL and Heartstart FR2+)
Materials
- Support: 1 millimeter thick white polyethylene foam and waterproof cell with biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic-based hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic PVC
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate/thermoplastic elastomer
Electrical properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS regulations
- Directive 93/42/EEC, Class I medical product