Pre-connected adult defibrillation electrodes for Agilent Philips equipment. 1 pair
Pair of preconnected disposable electrodes indicated for defibrillation, ECG monitoring and synchronized cardioversion. Compatible with Agilent Philips equipment.
Description
Pair of disposable, non-sterile, and pre-connected electrodes indicated for Agilent Philips Heartstart XL and Heartstart FR2+ equipment. Manufactured with biocompatible materials and designed for easy placement on the patient. Suitable for use in defibrillation procedures, synchronized cardioversion, or monitoring via electrocardiogram.
General Features
- Non-sterile disposable electrodes suitable for defibrillation, ECG monitoring, and synchronized cardioversion
- Supplied with the connection cable outside the bag to keep them pre-connected, ideal for emergencies
- Highly resistant, can withstand up to 15 discharges at 360 J sequentially
- Made without natural latex, phthalates, or biological origin compounds
- Non-cytotoxic, non-sensitizing, and non-irritating
Technical Specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the date of manufacture
- Compatibility: Agilent Philips equipment (Heartstart XL and Heartstart FR2+)
Materials
- Support: 1-millimeter thick white polyethylene flexible foam and impermeable cell with biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic base hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic polyvinyl chloride
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate/thermoplastic elastomer
Electrical Properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and Certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS Directive
- Directive 93/42/EEC, class I medical device
General Features
- Non-sterile disposable electrodes suitable for defibrillation, ECG monitoring, and synchronized cardioversion
- Supplied with the connection cable outside the bag to keep them pre-connected, ideal for emergencies
- Highly resistant, can withstand up to 15 discharges at 360 J sequentially
- Made without natural latex, phthalates, or biological origin compounds
- Non-cytotoxic, non-sensitizing, and non-irritating
Technical Specifications
- Electrode dimensions: 105 x 155 mm
- Active area dimensions: 79.45 cm2
- Recommended usage time: 24 h
- Minimum patient weight: 10 kg
- Shelf life: 2 years from the date of manufacture
- Compatibility: Agilent Philips equipment (Heartstart XL and Heartstart FR2+)
Materials
- Support: 1-millimeter thick white polyethylene flexible foam and impermeable cell with biocompatible acrylic adhesive
- Conductive area: tin foil with biocompatible acrylic base hydrogel
- Protector: polyethylene terephthalate
- Connection cable: copper (65/0.127 mm) coated with non-toxic polyvinyl chloride
- Connector mold: plastic material
- Ring terminals: brass
- Washer: ethylene vinyl acetate/thermoplastic elastomer
Electrical Properties
- Low frequency AC impedance at 10 Hz: < 3000 Ω
- Low frequency AC impedance at 30 kHz: < 10 Ω
- High frequency AC impedance: < 5 Ω
- Defibrillation overload recovery after 4 s: < 500 mV
- Defibrillation overload recovery after 60 s: < 400 mV
- DC Offset Voltage: < 100 mV
Standards and Certifications
- IEC 60601-2-4
- ISO 10993-5
- ISO 10993-10
- ISO 13485
- RoHS Directive
- Directive 93/42/EEC, class I medical device