Reference. 4047QE12-28SOB_24V

Emergency battery for QE12-28SOB surgical lamps of 24V

Emergency supply unit for ceiling surgical lamps compatible with Quattroluci Led, Pentaled 12, Pentaled 28 and Saturno-Led.
Personalized attention
Personal advice
Emergency battery for QE12-28SOB surgical lamps of 24V
Emergency battery for QE12-28SOB surgical lamps of 24V
1.030,00€ (1.246,30€ VAT INCLUDED)
Buy
Secure payment
Approximate delivery time of 10 to 20 days

Description

Rimsa
The QE12-28SOB 24V emergency power supply team allows for an automatic switch from the grid to the battery, ensuring that the lamp and its controls remain operational for a certain period of time, allowing an intervention to continue in the event of an unexpected power outage or power failure.

Technical specifications:
- Voltage: 12 V.
- Capacity 12 AH.
- Maximum charging current: 3A.
- Weight: 3.75 g.
- Dimensions: 151 x 98 x 94 (mm).
- Operating range: -20ºC / 40º.

Autonomy:
- Quattroluci Led: 2 h.
- Pentaled 12: 2 h.
- Pentaled 28: 1.5 h.
- Saturno-Led: 4 h.
- If the battery is completely discharged, the theoretical number of life cycles is 250.
- If the battery is discharged by 50% and then recharged, the theoretical number of life cycles is 550.
- If the battery is discharged by 30% and then recharged, the theoretical number of life cycles is 1200.

Recommendations:
- Perform a charge/discharge cycle of the battery once a month:
      · Charge the battery for at least 6 hours.
      · Leave the device on until the battery is exhausted.
      · If the battery's operating time mentioned in "Autonomy" is not reached, it should be replaced.
-To ensure good battery operation, replace it at least every 3 years.

First ignition:
The battery is shipped already charged and with its relative fuse (F3) disconnected and separate.
You will need to connect the battery using the faston and inserting the fuse.
Before connecting to the power system, verify that all fuses have been inserted.
Once turned on, check that the system operates correctly in relation to the charge and discharge cycle.

Recharge cycle:
The battery recharging cycle is indicated in color LEDs:
- Red: battery charging.
- Yellow: battery charged between 80% and 95%.
- Green: battery fully charged.
The LED light stays on when the battery is charging.