Reference. 94278VE12VET
Multiparameter monitor for veterinary VE12 (ECG, SpO2, NIBP, RESP, TEMP and PR)
Multiparameter monitor designed for veterinary diagnosis. Capable of accurately measuring all kinds of vital constants, including heart rate, blood pressure, respiration, body temperature, blood oxygen level, and heart electrical activity.
Personal advice
Description
This state-of-the-art equipment has been designed to perform vital sign monitoring in veterinary patients. Its operating system, reinforced with multiple features and algorithms, provides exceptional precision and reliability, even in situations where there are unstable connections or interrupted measurements.
The monitor also has a modular design intended to easily and conveniently incorporate new functionalities. This version allows readings of respiration, blood oxygen level, non-invasive blood pressure, heart rate, electrical activity of the heart, and body temperature. Optionally and at an additional cost, modules can be purchased to measure exhaled carbon dioxide and invasive blood pressure. Please contact us if you would like more information on this.
General Features
- High precision multiparameter monitor designed for animal diagnosis
- Modularly designed to accommodate additional monitoring systems
- ECG and SpO2 readings by fusion of homologous signals captured by intraoperative cable
- Highly accurate and reliable measurements, even in situations of instability or interruption of measurement
- Includes a rechargeable battery with 30 minutes of autonomy for emergency situations
- Incorporates multiple alarms to alert against abnormalities that may endanger the veterinary patient
General Technical Specifications
- Power supply: 220 V AC, 50/60 Hz (maximum current of 70 VA)
- Electrical protection standard: class I (with internal power)
- Electrical resistance standard: CF application
- Antiliquid level: equipment with standard sealing (without specific sealing for liquid prevention)
- Screen indications: up to 6 simultaneous waves, alarm indicator (yellow/red), operation indicator (green) and charging indicator (green). Three alarm tones.
- Battery autonomy: > 30 minutes
- Printer paper compatibility: 48 mm wide paper
- Printer speed: 25 mm/s
- Printer waves: 2
- Record type: 8 s real-time record
- Frequency in short sampling (1 hour): resolution every 1 or 5 seconds
- Frequency in long sampling (72 hours): 1, 5 or 10 minutes
- Storage: up to 400 PANI records
TEMP Module
- Sensor: YSI/CYF
- Channels: 2
- Measurement range: 0 - 50ºC
- Resolution: 0.1ºC
- Accuracy: ±0.1ºC (without error sensor included)
RESP Module
- Measurement method: RA-LL impedance method
- Range of respiratory impedance detection: 0.3 - 3 Ω
- Base impedance range: 200 - 400 Ω
- Bandwidth: 0.1 - 2.5 Hz
- Respiratory rate (dogs/cats): 0 - 150 rpm
- Resolution: 1 rpm
- Accuracy: ± 2 rpm
- Asphyxia alarm: 10 - 40 s
SpO2 Module
- Measurement mechanism with interference correction and weak perfusion correction
- SpO2 measurement range: 0 - 100%
- Resolution SpO2: 1%
- Accuracy SpO2 (dogs): 70 - 100%, ± 2; 0 - 69%, undefined
- Accuracy SpO2 (cats): 70 - 100%, ± 3; 0 - 69%, undefined
- Pulse rate measurement range: 20 - 300 bpm
- Pulse rate resolution: 1 bpm
- Pulse rate accuracy: ± 3 bpm
- Pulse rate data update: 1 s
NIBP (Non-Invasive Blood Pressure) Module
- Measurement system designed for animal cardiac characteristics
- Reinforced with algorithms to correct interference and weak signals
- Measurement method: pulse wave oscillation
- Measurement model: manual/automatic
- Measurement intervals in automatic mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 min
- Measurement time in continuous measurement: 5 min
- Heart rate range: 40 - 240 bpm
- Contraction pressure range: 40 - 270 mmHg
- Diastolic blood pressure: 10 - 215 mmHg
- Average pressure: 20 - 235 mmHg
- Static pressure range: 0 - 300 mmHg
- Static pressure accuracy: ±3 mmHg
- Blood pressure accuracy: ±5 mmHg (maximum average error), 8 mmHg (maximum standard deviation)
ECG and PR Modules
- Latest generation ECG measurement algorithm patented by Mindray
- Capable of detecting a wide spectrum of arrhythmic dysfunctions
- Measurement of the average heart rate and reading sweep every second
- Can perform readings in patients with and without pacemakers
- Lead configuration: 3 leads; RA, LA, LL; lead mode: I, I, I I
- Acquisition: automatic, 0.25, 0.5 or 1.2
- Deviation: < 0.66% per minute, < ±10% per hour
- Heart rate measurement range: 15 - 350 bpm (dogs/cats)
- Selection and accuracy of the time reference point: 25 mm/s (6.25 mm/s, 2.5 mm/s and 50 mm/s optional); duration of error < ±10%
- Response time to changes in heart rate: increase 80 - 120 bpm in less than 12 s; decrease from 80 - 40 bpm in less than 12 s
- Alarm activation time: cardiac arrest in less than 12 s; low heart rate in less than 12 s and high heart rate in less than 12 s
- Heart rate alarm range (dogs/cats): 17 - 350 bpm (upper limit) and 15 - 348 bpm (lower limit), resolution of 1 bpm and error limit of ±1% or ±1 bpm
- Interference suppression: variations of < ±10% due to interference suppression
- Overload protection: > 1 V
- QRS wave range (P-V RTI): 0.5 - 5 mV
- QRS wave width (dogs/cats): 40 - 120 ms
- Lack of response: range (P-V RTI) < 1.5 mV, wave width of 1 mV < 10 ms
- Power frequency voltage tolerance: > 100 uV
- Input dynamic range: wave amplitude < ± 5 mV; rhythm (RTI) < 320 mV/s; DC tendency voltage -300 - 300 mV; output signal change < ±10%; screen failure < 50%
- System noise: < 30 μV
- Multichannel cross-communication: < 5%
- Channel amplitude: > 30 mm, aspect ratio of 0.4 s/mV
- Input signal reconstruction accuracy: total error ±20% or greater than 100% μV
- Frequency response: input sinusoidal frequency of 0.67 - 40 Hz, corresponding to output frequency of 0.67 - 40 Hz (attenuation of -3dB). The 20 ms input triangular wave attenuates 0 - 20% against the highest amplitude.
- Common mode rejection: >90 dB (diagnostic mode), >100 dB (monitoring mode) and >100 dB (operational standard)
- Electrode polarization voltage range: ± 300 mV
- Suppression of fast ECG signals by pacemaker with pulse detector: 1 V/s RTI, 5% accuracy
- Stability and control of reference point: recovery time < 3 s after reset; drift rate < 10 s than 10 uV/s; reference point drift of < 500 uV in 1 h and reference point drift < 50 uV/C at room temperature
- Drift tolerance: triangular amplitude (P-V RTI) of 4 mV, QRS amplitude (P-V RTI) of 0.5 mV, QRS wave amplitude of 100 ms, QRS wave repetition of 80 bpm
Optional Accessories (not included and at an additional cost)
- EtCO2 Module (Ref. 94278VE12VET-ETCO2)
- Nellcor blood oxygen module
- SunTech blood pressure module
- IAP (Invasive Arterial Pressure) module
The monitor also has a modular design intended to easily and conveniently incorporate new functionalities. This version allows readings of respiration, blood oxygen level, non-invasive blood pressure, heart rate, electrical activity of the heart, and body temperature. Optionally and at an additional cost, modules can be purchased to measure exhaled carbon dioxide and invasive blood pressure. Please contact us if you would like more information on this.
General Features
- High precision multiparameter monitor designed for animal diagnosis
- Modularly designed to accommodate additional monitoring systems
- ECG and SpO2 readings by fusion of homologous signals captured by intraoperative cable
- Highly accurate and reliable measurements, even in situations of instability or interruption of measurement
- Includes a rechargeable battery with 30 minutes of autonomy for emergency situations
- Incorporates multiple alarms to alert against abnormalities that may endanger the veterinary patient
General Technical Specifications
- Power supply: 220 V AC, 50/60 Hz (maximum current of 70 VA)
- Electrical protection standard: class I (with internal power)
- Electrical resistance standard: CF application
- Antiliquid level: equipment with standard sealing (without specific sealing for liquid prevention)
- Screen indications: up to 6 simultaneous waves, alarm indicator (yellow/red), operation indicator (green) and charging indicator (green). Three alarm tones.
- Battery autonomy: > 30 minutes
- Printer paper compatibility: 48 mm wide paper
- Printer speed: 25 mm/s
- Printer waves: 2
- Record type: 8 s real-time record
- Frequency in short sampling (1 hour): resolution every 1 or 5 seconds
- Frequency in long sampling (72 hours): 1, 5 or 10 minutes
- Storage: up to 400 PANI records
TEMP Module
- Sensor: YSI/CYF
- Channels: 2
- Measurement range: 0 - 50ºC
- Resolution: 0.1ºC
- Accuracy: ±0.1ºC (without error sensor included)
RESP Module
- Measurement method: RA-LL impedance method
- Range of respiratory impedance detection: 0.3 - 3 Ω
- Base impedance range: 200 - 400 Ω
- Bandwidth: 0.1 - 2.5 Hz
- Respiratory rate (dogs/cats): 0 - 150 rpm
- Resolution: 1 rpm
- Accuracy: ± 2 rpm
- Asphyxia alarm: 10 - 40 s
SpO2 Module
- Measurement mechanism with interference correction and weak perfusion correction
- SpO2 measurement range: 0 - 100%
- Resolution SpO2: 1%
- Accuracy SpO2 (dogs): 70 - 100%, ± 2; 0 - 69%, undefined
- Accuracy SpO2 (cats): 70 - 100%, ± 3; 0 - 69%, undefined
- Pulse rate measurement range: 20 - 300 bpm
- Pulse rate resolution: 1 bpm
- Pulse rate accuracy: ± 3 bpm
- Pulse rate data update: 1 s
NIBP (Non-Invasive Blood Pressure) Module
- Measurement system designed for animal cardiac characteristics
- Reinforced with algorithms to correct interference and weak signals
- Measurement method: pulse wave oscillation
- Measurement model: manual/automatic
- Measurement intervals in automatic mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 min
- Measurement time in continuous measurement: 5 min
- Heart rate range: 40 - 240 bpm
- Contraction pressure range: 40 - 270 mmHg
- Diastolic blood pressure: 10 - 215 mmHg
- Average pressure: 20 - 235 mmHg
- Static pressure range: 0 - 300 mmHg
- Static pressure accuracy: ±3 mmHg
- Blood pressure accuracy: ±5 mmHg (maximum average error), 8 mmHg (maximum standard deviation)
ECG and PR Modules
- Latest generation ECG measurement algorithm patented by Mindray
- Capable of detecting a wide spectrum of arrhythmic dysfunctions
- Measurement of the average heart rate and reading sweep every second
- Can perform readings in patients with and without pacemakers
- Lead configuration: 3 leads; RA, LA, LL; lead mode: I, I, I I
- Acquisition: automatic, 0.25, 0.5 or 1.2
- Deviation: < 0.66% per minute, < ±10% per hour
- Heart rate measurement range: 15 - 350 bpm (dogs/cats)
- Selection and accuracy of the time reference point: 25 mm/s (6.25 mm/s, 2.5 mm/s and 50 mm/s optional); duration of error < ±10%
- Response time to changes in heart rate: increase 80 - 120 bpm in less than 12 s; decrease from 80 - 40 bpm in less than 12 s
- Alarm activation time: cardiac arrest in less than 12 s; low heart rate in less than 12 s and high heart rate in less than 12 s
- Heart rate alarm range (dogs/cats): 17 - 350 bpm (upper limit) and 15 - 348 bpm (lower limit), resolution of 1 bpm and error limit of ±1% or ±1 bpm
- Interference suppression: variations of < ±10% due to interference suppression
- Overload protection: > 1 V
- QRS wave range (P-V RTI): 0.5 - 5 mV
- QRS wave width (dogs/cats): 40 - 120 ms
- Lack of response: range (P-V RTI) < 1.5 mV, wave width of 1 mV < 10 ms
- Power frequency voltage tolerance: > 100 uV
- Input dynamic range: wave amplitude < ± 5 mV; rhythm (RTI) < 320 mV/s; DC tendency voltage -300 - 300 mV; output signal change < ±10%; screen failure < 50%
- System noise: < 30 μV
- Multichannel cross-communication: < 5%
- Channel amplitude: > 30 mm, aspect ratio of 0.4 s/mV
- Input signal reconstruction accuracy: total error ±20% or greater than 100% μV
- Frequency response: input sinusoidal frequency of 0.67 - 40 Hz, corresponding to output frequency of 0.67 - 40 Hz (attenuation of -3dB). The 20 ms input triangular wave attenuates 0 - 20% against the highest amplitude.
- Common mode rejection: >90 dB (diagnostic mode), >100 dB (monitoring mode) and >100 dB (operational standard)
- Electrode polarization voltage range: ± 300 mV
- Suppression of fast ECG signals by pacemaker with pulse detector: 1 V/s RTI, 5% accuracy
- Stability and control of reference point: recovery time < 3 s after reset; drift rate < 10 s than 10 uV/s; reference point drift of < 500 uV in 1 h and reference point drift < 50 uV/C at room temperature
- Drift tolerance: triangular amplitude (P-V RTI) of 4 mV, QRS amplitude (P-V RTI) of 0.5 mV, QRS wave amplitude of 100 ms, QRS wave repetition of 80 bpm
Optional Accessories (not included and at an additional cost)
- EtCO2 Module (Ref. 94278VE12VET-ETCO2)
- Nellcor blood oxygen module
- SunTech blood pressure module
- IAP (Invasive Arterial Pressure) module