Autoclave for sterilization Autester ST Dry PV III, 50 liters
Autester ST Dry PV III Autoclave for Sterilization, 50 liters
MICROPROCESSOR PROCESS CONTROL. AUTOMATIC DRYING AND PURGING SYSTEM BY FRACTIONAL VACUUM. FOR ADJUSTABLE TEMPERATURES FROM 105 °C TO 134 °C (0.21 TO 2 bar). COMPLIES WITH PRESSURE EQUIPMENT DIRECTIVES. VERTICAL MODEL
FEATURES
Exterior...
Description
Autester ST Dry PV III Autoclave for Sterilization, 50 liters
MICROPROCESSOR PROCESS CONTROL. AUTOMATIC DRYING AND PURGING SYSTEM BY FRACTIONAL VACUUM. FOR ADJUSTABLE TEMPERATURES FROM 105 °C TO 134 °C (0.21 TO 2 bar). COMPLIES WITH PRESSURE EQUIPMENT DIRECTIVES. VERTICAL MODEL
FEATURES
Exterior cabinet and countertop in stainless steel.
Tank and lid in stainless steel.
Purge by fractional vacuum and atmospheric pump.
External steam generator. Water tank to feed the boiler.
Possibility of feeding the boiler from an external water intake. 6 operating modes.
Output for printer. RS 232 output for printing parameters from a computer.
USB output for data logging on PEN-DRIVE and connection to computer.
Connection for a second control probe.
Storage for 9 programs totally customizable by the user.
SAFETY
Device that prevents the lid from opening while the chamber is under pressure.
Safety valve. Prevents the pressure from exceeding the maximum limit.
Safety pressure switch. Disconnects the heating in case of overpressure.
Thermal protection casing for the lid.
Correctly closed door detector.
MICROPROCESSOR FUNCTIONS
The microprocessor controls the completion of the air extraction, sterilization and drying process by executing the procedures and orders selected by the operator, regulating the following parameters:
• Sterilization temperature from 105 °C to 134 °C.
• Sterilization time from 3' to 59'.
• Drying time from 20' to 60'.
• Up to 10 programs customizable by the user.
• Six operating modes: - Solids 1 - Solids 2 fractions. - Solids 3 fractions. - Drying. - Liquids. - Liquids 2 probes.
• Verification test - Vacuum. - Bowie Dick.
• Configuration - Units. - Languages (Spanish, English, French). - USB module or printer. - Clock: date/time. - External water intake: on/off.
• Password
• Calibration - Probe 1 temperature - Probe 2 temperature - Water dosing.
• Audible and visual over temperature and insufficient temperature alarm.
• Detection and memorization of network failures
• Software safety thermostat (TS)
• Sterilization cycle data logging with capacity for more than 500 cycles
• Internal cleaning program
• PC software
• Temperature correction
• Abbreviated user manual on screen
• Real-time graphical representation of the process