Autoclave 12 liters class B IcanClave
Description
Buy Autoclave 12 liters class B
Autoclave 12 liters class B with USB and double safety lock.Technical specifications
- Nominal Voltage: 220-240 VAC; 50/60 Hz; 10 A.
- Nominal Power: 2000 W.
- Circuit Breaker: F16A / 400 V.
- Chamber dimensions: 200 x 360mm.
- Sterilization temperature: 121 °C / 134 °C.
- Distilled water tank capacity: 2.5L.
- Operating temperature: 5 °C / 40 °C.
- Operating relative humidity: Max. 80%, non-condensing.
- External dimensions: 414 x 365 x 570 mm.
- Net Weight: 36.5 kg.
- Operating atmospheric pressure: 76kPa / 106 kPa.
- Maximum noise level: <70 dB.
Components
- Door with automatic locking system.
- Opening and closing handle.
- LCD screen.
- Control panel.
- Distilled water tank.
- USB port.
- Key for draining distilled and used water.
- Safety valve.
- Ventilation grid.
- Power cord.
Included accessories
- 3 Stainless steel instrument trays.
- Tray base.
- Tray extraction handle.
- 2 Drain hoses.
- Door adjustment tool.
- Instruction manual.
Installation
- The sterilizer should be placed on a flat, level surface with a minimum load capacity of 60 Kg.
- To ensure proper air circulation, a free space of at least 10cm on the sides and 20cm on the top should be left around the device.
- The sterilizer should be located in a properly ventilated enclosure.
- Do not install the sterilizer near sources of water and/or heat.
- Do not cover the condenser fan's air outlet grille, nor block the side outlets.
Regulation
This Icanclave sterilizer has been designed and manufactured in accordance with the following directives and standards:
- Directive 97/23 / EC on pressure equipment.
- Directive 93/42 / EEC on class II-a medical devices.
- EN13060 standard for small steam sterilizers.
- EN61010-1 safety standard for laboratory equipment.
- EN 61010-2-040 safety standard for medical material sterilizers.
- EN 61326-1 electromagnetic compatibility standard.