Metallic container 20x10x6 cm. Autoclavable
Description
Stainless steel box without handles
Dimensions: 20 x 10 x 6 cm
Autoclavable 100 ° C
The product must be thoroughly cleaned with hot water for the first time and after each use, disinfected, possibly rinsed with distilled water, and dried with extreme care. It must also be sterilized before subsequent use.
The product can be cleaned in ultrasonic tanks provided for this purpose and disinfected in instrument washing machines up to a temperature of 95 ° C (203 ° F). After cleaning, and before sterilization, it is recommended to treat the instruments with physiologically safe oil, especially with regard to the tips, connectors, terminals and all moving parts. Great care must be taken to ensure that the product does not come into contact with acids or other aggressive disinfectants that could corrode it.
The recommended sterilization method is steam in an autoclave, the cycle temperature should not exceed 135 ° C (275 ° F) for a maximum of 15 minutes, to avoid damaging the product.
For hot air sterilization, a temperature between 180 ° C and 200 ° C is advised.
Products can also be disinfected in instrument washing machines up to a temperature of 121 ° C.
Instruments showing signs of corrosion must be immediately removed.